These instructions can help you learn how to use the AeroChamber Plus® Flow-Vu® with a Metered Dose Inhaler (MDI). Watch the video or read the instructions below. Print the full AeroChamber Plus Flow-Vu aVHC Instructions for Use in English and Spanish by clicking the button below: 

THIS PRODUCT CAN BE USED RIGHT OUT OF PACKAGE. BEFORE USE, ENSURE THESE INSTRUCTIONS AND THE INSTRUCTIONS SUPPLIED WITH THE METERED DOSE INHALER (MDI) HAVE BEEN READ.

 


Carefully examine the product for damage, missing parts, or foreign objects. Remove any foreign objects prior to use. The product should be replaced IMMEDIATELY if there are any damaged or missing parts. If necessary, use the Metered Dose Inhaler (MDI) alone until a replacement is obtained. If the patient's symptoms worsen, please seek immediate medical attention.

 


Remove cap from the MDI.

 


Shake the MDI immediately before each use as per the instructions supplied with the MDI.

 


Insert the MDI into the backpiece of the chamber.

 


Apply mask to face and ensure an effective seal. The Flow-Vu® Inspiratory Flow Indicator (IFI) only moves if a good seal is created.

 
 


Press the MDI at the beginning of a slow inhalation as the Flow-Vu® Inspiratory Flow Indicator moves toward the patient. Use the Flow-Vu® Inspiratory Flow Indicator to count breaths. Maintain seal for 5-6 breaths after the MDI is pressed. Administer one (1) puff at a time. 


Follow instructions supplied with the MDI on how long to wait before repeating steps 3-6 as prescribed.

Helpful Tips

  • Give praise and rewards.
  • Use teddy bears or dolls to 'pretend' to use the chamber to make it look like a game.
  • If possible, give treatments when the child is happy and not crying.
 
Downloads
 


Carefully examine the product for damage, missing parts or foreign objects. Remove any foreign objects prior to use. The product should be replaced IMMEDIATELY if there are any damaged or missing parts. If necessary, use the Metered Dose Inhaler (MDI) alone until a replacement is obtained. If symptoms worsen, please seek immediate medical attention.

 


Remove caps from the MDI and chamber.

 


Shake the MDI immediately before each use as per the instructions supplied with the MDI.

 


Insert the MDI into the backpiece of the chamber.

 


 

Put mouthpiece into mouth
and close lips around it to
ensure an effective seal. The
Flow-Vu® Inspiratory Flow Indicator (IFI) only moves if the patient has a good seal.


Apply mask to face
and ensure an effective
seal. The Flow-Vu®
Inspiratory Flow Indicator only moves if
a good seal is created.

 

 


 

Breathe out gently and press the MDI at the beginning of a slow inhalation. Use the Flow-Vu® IFI to assist in the coordination of this step. Breathe in slowly and deeply through the mouth until a full breath has been taken. Hold breath for 5-10 seconds, if possible. Otherwise, keep lips tight on the mouthpiece breathing normally 2-3 times through the chamber after the MDI is pressed. Slow down inhalation if you hear the FlowSignal® Whistle sound. It means that you are inhaling too quickly. Administer one (1) puff at a time. 



Breathe out gently and press
the MDI at the beginning of
a slow inhalation as the
Flow-Vu® IFI moves toward
you. Use the Flow-Vu® IFI to
count breaths. Maintain seal for
5-6 breaths after the MDI is pressed. Slow down inhalation if you hear the FlowSignal® Whistle sound.
It means that you are inhaling too quickly. Administer one (1) puff at a time. 

 

 


Follow instructions supplied with the MDI on how long to wait before repeating steps 3-6 as prescribed.

Downloads
INDICATIONS FOR USE AND CAUTIONS

This product is intended to be used by patients who are under the care or treatment of a physician or licensed healthcare professional. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers. The intended environments for use include the home, hospitals and clinics.

CAUTIONS:
  • PRODUCT MAY BE PERMANENTLY DAMAGED IF BOILED, STERILIZED,
    OR CLEANED IN A DISHWASHER AT TEMPERATURES ABOVE 158°F (70°C).
  • Do not leave the chamber unattended with children.
  • Federal (USA) law restricts the sale of this device on or by the order of a physician.