AeroChamber Plus® Flow-Vu® is dishwasher safe on the top rack at temperatures up to 158°F (70°C).

Learning How to Use Your AeroChamber Plus® Flow-Vu®

Watch the video below for instructions on using your AeroChamber Plus® with a metered dose inhaler, and for cleaning directions.

Download Instructions for Use—available in English and Spanish and featuring the child-friendly AeroBear®1 Or simply read the directions by clicking the tabs below.

Aero Bearsmall-medium-mask-1large-mask-1
Corner

tab1-step1

Carefully examine the product for damage, missing parts, or foreign objects. Remove any foreign objects prior to use. The product should be replaced IMMEDIATELY if there are any damaged or missing parts. If necessary, use the Metered Dose Inhaler (MDI) alone until a replacement is obtained. If the patient's symptoms worsen, please seek immediate medical attention.

tab1-step2

Remove cap from the MDI.

tab1-step3

Shake the MDI immediately before each use as per the instructions supplied with the MDI.

tab1-step4

Insert the MDI into the backpiece of the chamber.

tab1-step5

Apply mask to face and ensure an effective seal. The Flow-Vu® Inspiratory Flow Indicator (IFI) only moves if a good seal is created.

tab1-step6

Press the MDI at the beginning of a slow inhalation as the Flow-Vu® Inspiratory Flow Indicator moves toward the patient. Use the Flow-Vu® Inspiratory Flow Indicator to count breaths. Maintain seal for 5-6 breaths after the MDI is pressed. Administer 1 puff at a time.

tab1-step7

Follow instructions supplied with the MDI on how long to wait before repeating steps 3-6 as prescribed.

Helpful Tips

  • Give praise and rewards.
  • Use teddy bears or dolls to "pretend" to use the chamber to make it look like a game.
  • If possible, give treatments when the child is happy and not crying.

Downloads

Corner

CLEANING INSTRUCTIONS FOR THE SMALL AND MEDIUM MASK CHAMBERS. THIS PRODUCT CAN BE USED RIGHT OUT OF THE PACKAGE AND THEN CLEANED WEEKLY.

tab2-step1

Remove the backpiece. To detach the frontpiece, twist chamber as shown.

tab2-step2a

Soak the parts for 15 minutes in a mild solution of liquid dish detergent and lukewarm clean water. Agitate gently.

Rinse parts in clean water.

OR

tab2-step2b

Dishwasher safe under the following conditions:

  • Avoid heavy or other cycles that include heated dry over 158°F
  • Parts must be placed in the top rack ONLY 
  • Secure product face up, as pictured
  • Use dishwasher detergent and rinse aid
  • Do not boil or sterilize

SEE CAUTIONS, 1st BULLET, BELOW

tab2-step3

Shake out excess water from the parts and allow to air dry in a vertical position. Ensure parts are dry before reassembly.

tab2-step4

To reassemble, fit the frontpiece on the end of the chamber and twist firmly until securely locked into position.

tab2-step5

Center the Alignment Feature on the backpiece with the Flow-Vu® Inspiratory Flow Indicator, as shown. Press firmly to attach the backpiece.

 blue

tab3-step1

Carefully examine the product for damage, missing parts or foreign objects. Remove any foreign objects prior to use. The product should be replaced IMMEDIATELY if there are any damaged or missing parts. If necessary, use the Metered Dose Inhaler (MDI) alone until a replacement is obtained. If symptoms worsen, please seek immediate medical attention.

tab3-step2

Remove caps from the MDI and chamber.

tab3-step3

Shake the MDI immediately before each use as per the instructions supplied with the MDI.

tab3-step4

Insert the MDI into the backpiece of the chamber.

tab3-step5a

Put mouthpiece into mouth and close lips around it to ensure an effective seal.

The Flow-Vu® Inspiratory Flow Indicator (IFI) only moves if the patient has a good seal.

instructions-for-use-adults-5-2

Apply mask to face and ensure an effective seal. The Flow-Vu® Inspiratory Flow Indicator only moves if a good seal is created.

instructions-for-use-adults-5-3
tab3-step6a

Breathe out gently and press the MDI at the beginning of a slow inhalation. Use the Flow-Vu® IFI to assist in the coordination of this step. Breathe in slowly and deeply through the mouth until a full breath has been taken. Hold breath for 5-10 seconds, if possible. Otherwise, keep lips tight on the mouthpiece breathing normally 2-3 times through the chamber after the MDI is pressed. Slow down inhalation if you hear the FlowSignal® Whistle sound. It means that you are inhaling too quickly. Administer 1 puff at a time.

Breathe out gently and press the MDI at the beginning of a slow inhalation as the Flow-Vu® IFI moves toward you. Use the Flow-Vu® IFI to count breaths. Maintain seal for 5-6 breaths after the MDI is pressed. Slow down inhalation if you hear the FlowSignal® Whistle sound. It means that you are inhaling too quickly. Administer 1 puff at a time.

Instructions For Use Adults 6

Breathe out gently and press the MDI at the beginning of a slow inhalation as the Flow-Vu® IFI moves toward you. Use the Flow-Vu® IFI to count breaths. Maintain seal for 5-6 breaths after the MDI is pressed.Slow down inhalation if you hear the FlowSignal® Whistle sound. It means that you are inhaling too quickly. Administer 1 puff at a time.

Instructions For Use Adults 6
tab3-step7

Follow instructions supplied with the MDI on how long to wait before repeating steps 3-6 as prescribed.

Downloads

 blue

CLEANING INSTRUCTIONS FOR THE LARGE MASK AND MOUTHPIECE CHAMBERS. THIS PRODUCT CAN BE USED RIGHT OUT OF THE PACKAGE AND THEN CLEANED WEEKLY.

tab4-step1

Remove the backpiece. To detach the frontpiece, twist chamber as shown. Remove the mouthpiece cap.

tab4-step2

Soak the parts for 15 minutes in a mild solution of liquid dish detergent and lukewarm clean water. Agitate gently.

Rinse parts in clean water.

OR

alt

Dishwasher safe under the following conditions:

  • Avoid heavy or other cycles that include heated dry over 158°F
  • Parts must be placed in the top rack ONLY 
  • Secure product face up, as pictured
  • Use dishwasher detergent and rinse aid
  • Do not boil or sterilize

SEE CAUTIONS, 1st BULLET, BELOW

 

tab4-step3

Shake out excess water from the parts and allow to air dry in a vertical position. Ensure parts are dry before reassembly.

tab4-step4

To reassemble, fit the frontpiece on the end of the chamber and twist firmly until securely locked into position.

For mouthpiece models, the protective cap should always be placed on the mouthpiece when the chamber is not in use.

tab4-step5

Center the Alignment Feature on the backpiece with the Flow-Vu® Inspiratory Flow Indicator, as shown. Press firmly to attach the backpiece.

Notes:

  • Storage and operating range is 5°C–40°C (41°F–104°F) at 15% to 95% relative humidity
  • Product may need to be replaced after 12 months of use. Environmental conditions, storage and proper cleaning can affect
    product life span
  • Do not share this medical device
  • If you notice medication build-up in your chamber, wash the inside of the chamber gently with a soft cloth
  • Dishwashing with overly dirty dishes is not recommended
  • Dishwasher validated up to 158°F (70°C)

Bpa Safe

AeroChamber Plus® Flow-Vu® is dishwasher safe on the top rack at temperatures up to 158°F (70°C).

INDICATIONS FOR USE

This product is intended to be used by patients who are under the care or treatment of a physician or licensed healthcare professional. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers. The intended environments for use include the home, hospitals and clinics.

CAUTIONS:
  • PRODUCT MAY BE PERMANENTLY DAMAGED IF BOILED, STERILIZED, OR CLEANED IN A DISHWASHER AT TEMPERATURES ABOVE 158°F (70°C).
  • Do not leave the chamber unattended with children.
  • Federal (USA) law restricts the sale of this device on or by the order of a physician.

Please see Instructions for Use.

Available by prescription only.

This product is intended to be used by patients who are under the care or treatment of a physician or licensed healthcare professional. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers. The intended environments for use include the home, hospitals and clinics.

CAUTIONS:
  • PRODUCT MAY BE PERMANENTLY DAMAGED IF BOILED, STERILIZED, OR CLEANED IN A DISHWASHER AT TEMPERATURES ABOVE 158°F (70°C).
  • Do not leave the chamber unattended with children.
  • Federal (USA) law restricts the sale of this device on or by the order of a physician.

Please see Instructions for Use.

Available by prescription only.